HARDWARE

ADWin Based DYNTTS/EMDCSS/HCA

Hardware Troubleshooting Guide

Unable to render {include} The included page could not be found.

The purpose of this helpsheet is to help to diagnose hardware problems with ADWin based dynamic systems - DYNTTS, EMDCSS, HCA

ADWin DCS Control Box

There are two versions of the ADWin based control box, one based on the ADWin-Light16 processor and one based on the ADWin Gold processor.

Voltage Check:

Pin

Colour

Voltage

6

Blue

-15V

5

Black

0V

4

Black

0V

3

Red

+5V

1

Violet

+15V

Colour

Voltage

Blue

-15V

Black/braid

0V

Violet

+15V

DCS - 8 channel using ADwin-Light processor

Check that all connectors are firmly plugged in.

ADV-DCS - 8 or 16 channel using ADWin Gold processor

Common HARDWARE Diagnosis MINIDYN/EMDCSS/MAXIDYN/HCA

Voltage check:

Digital Input/Outputs

The simplest place to start diagnosing the base functionality of a GDS ADWin based Dynamic System is to use the ADWin Test v1.2 software, which is a piece of GDS debugging software, in conjunction with a good quality Multimeter to measure the Control voltages and the Digital Input and Output signal voltages.

GDS ADWin Test v1.2 allow debuggging of all of the data acquisition channel values, encoder values, VCS control Signals, Digital Outputs and Digital Inputs.

Open GDS_ADwin_Test.exe and select the relevant Tab for your Dynamic System (MINIDYN / EMDCSS / MAXIDYN/HCA).

MINIDYN

EMDCSS

MAXIDYN/HCA

Export the finished helpsheet as PDF to M:\Technical\Handbooks\GDSK Confluence Helpsheets\

Delete this Warning Panel when done