Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

HARDWARE

ADWin Based DYNTTS/EMDCSS/HCA

Hardware Troubleshooting Guide

Include Page
GIHS:GDS Logo
GIHS:GDS Logo

Info

The purpose of this helpsheet is to help to diagnose hardware problems with ADWin based dynamic systems - DYNTTS, EMDCSS, HCA

ADWin DCS Control Box

There are two versions of the ADWin based control box, one based on the ADWin-Light16 processor and one based on the ADWin Gold processor.

Voltage Check:

Image RemovedImage Added

Pin

Colour

Voltage

6

Blue

-15V

5

Black

0V

4

Black

0V

3

Red

+5V

1

Violet

+15V

Colour

Voltage

Blue

-15V

Black/braid

0V

Violet

+15V

DCS - 8 channel using ADwin-Light processor

Check that all connectors are firmly plugged in.

Image RemovedImage Added

ADV-DCS - 8 or 16 channel using ADWin Gold processor

Common HARDWARE Diagnosis MINIDYN/EMDCSS/MAXIDYN/HCA

Voltage check:

Digital Input/Outputs

The simplest place to start diagnosing the base functionality of a GDS ADWin based Dynamic System is to use the ADWin Test v1.2 software, which is a piece of GDS debugging software, in conjunction with a good quality Multimeter to measure the Control voltages and the Digital Input and Output signal voltages.

Image Removed

.

GDS ADWin Test v1.2 allow debuggging of all of the data acquisition channel values, encoder values, VCS control Signals, Digital Outputs and Digital Inputs.

Open GDS_ADwin_Test.exe and select the relevant Tab for your Dynamic System (MINIDYN / EMDCSS / MAXIDYN/HCA).

Image Added

MINIDYN

EMDCSS

MAXIDYN/HCA

Note

Export the finished helpsheet as PDF to M:\Technical\Handbooks\GDSK Confluence Helpsheets\

Delete this Warning Panel when done